物流技术网

登录

物流相关设备消毒设备灭菌柜

VHP灭菌柜

供应商:
苏州宝康机械科技有限公司
企业类型:
其他

产品简介

产品详情技术参数:NOTechnicalrequirementTechnicalparameterRemarks1设备内腔表面粗糙度/SurfaceroughnessoftheinnercavityofequipmentRa≤0

详细信息

产品详情

技术参数:

NO

Technical requirement

Technical parameter

Remarks

1

设备内腔表面粗糙度/Surface roughness of the inner cavity of equipment

Ra0.5um

 

2

无菌传递总时间/Total time of aseptic transfer

100-120min

The full load

3

除湿总时间/Total dehumidification time

30-45min

 

4

灭菌总时间/Total sterilization time

30-45min

 

5

通风总时间/Total ventilation time

60-90min

 

6

灭菌浓度/Sterilizing concentration

150-400PPM

 

7

灭菌后残留VHP浓度标准/Standard of residual VHP concentration after sterilization

1ppm

 

8

内腔材质/Inner cavity material

SUS304/T2.0mm

 

9

设备外表面材质/External surface material of equipment

SUS304

 

10

换气次数/Air change times

150-200pcs

 

11

设备运行阶段内部洁净等级/Internal cleanliness grade of equipment operation stage

A

 

12

设备泄露率/Equipment leakage rate

0.5%

 

13

噪声级别/Noise level

75db

 

14

灭菌效果/Sterilization effect

6LOG

 

15

总功率(W)/Total power (W)

≤3500W

 

16

非金属材料/Nonmetallic materials

PTFE)、Silica gelTempered glass

 

17

汽化温度/Vaporization temperature

100

 

18

重量/Weight

600KG

 

19

体积/volume

见方案

 

工艺描述

过氧化氢传递窗用于无菌物料由C/D级洁净区向A/B级洁净区传递时,去除物品或容器表面的生物污染。共分三个步骤:

步骤1物品或容器在C/D级区域放到传递窗内(放置时物品之间留有间隙,保证物品或容器限度的暴露在空气中,不留灭菌死角),关闭窗门后进行下一步。

步骤2关好窗门后接通电源,开启仪器进行灭菌,可选用手动和自动运行程序两种模式,手动模式下分为除湿、调节、灭菌、排残、通风、保压六个步骤其中汽化温度设定为95℃左右,调节周期和灭菌周期可根据实际情况进行设定,排残时间或通风时间可设置,调节间歇和灭菌间歇设置可设置,保压压差设定范围10pa-60pa在灭菌过程中,传递窗内的过氧化氢浓度传感器可实时监控窗内过氧化氢浓度。

步骤3灭菌结束后,打开A/B级洁净区的传递门,从传递窗内取出被灭物品